Issue Alert Help & FAQ Search Current Issue
ENTH Navbar



[Full-text Article] [PDF Version] [Create Reference]
[PubMed Citation] [Related Articles in PubMed]

Abstract


doi: 10.1583/04-1253MR.1
Journal of Endovascular Therapy: Vol. 11, No. 6, pp. 641–648.

Endovascular Repair of Abdominal Aortic Aneurysms Using 3 Commercially Available Devices: Midterm Results

Ali F. AbuRahma, MD1; Patrick A. Stone, MD1; Mark C. Bates, MD1; Tariq N. Khan, MD1; Jason M. Prigozen, MD1; and Christine A. Welch, MS1

1Vascular Center of Excellence and Department of Surgery, Robert C. Byrd Health Sciences Center of West Virginia University, Charleston, West Virginia, USA



Purpose: To report the early and midterm results of 3 commercially available stent-grafts for abdominal aortic aneurysm (AAA) exclusion.

Methods: Since fall 1999, 151 patients (137 men; mean age 74 years, range 54–88) have undergone AAA repair using 88 Ancure, 46 AneuRx, and 17 Excluder stent-grafts in elective procedures. All patients were followed clinically and underwent postoperative duplex ultrasound and/or computed tomographic angiography, which was repeated every 6 months.

Results: The mean overall follow-up was 17 months (range 1–46). Initial technical failure was 3% (3/88) for Ancure versus 0% for AneuRx and Excluder. Primary endoleak occurred in 8% (7/88) of the Ancure patients versus 22% (10/46) for AneuRx and 6% (1/17) for Excluder (p=NS). Early (30-day) graft thrombosis (2/151, 1%) was seen only in the Ancure group. The perioperative complication rates (excluding endoleak and fever) were 22% (19/ 88) for Ancure, 15% (7/46) for AneuRx, and 0% for Excluder (p=NS). There was only 1 (0.7%) perioperative death in the series (Ancure patient). More ancillary procedures were performed in the Ancure group (p<0.05). Postoperatively, the size of the AAA decreased or remained unchanged in 76% for Ancure patients versus 75% for AneuRx and 64% for Excluder (p=NS). The incidences of late endoleak were 6% for Ancure, 2% for AneuRx, and 0% for Excluder (p=NS). The freedom from late endoleak at 3 years was 88% for Ancure and 97% for AneuRx (100% at 1 year for Excluder). Survival rates were similar. No AAA rupture was recorded.

Conclusions: The Ancure device has slightly higher overall perioperative complications and early technical failure than AneuRx or Excluder. All 3 devices are effective in preventing aneurysm rupture; the overall midterm survival rates, freedom of endoleak, and limb patency are somewhat similar.

Keywords: abdominal aortic aneurysm, endovascular repair, stent-graft, outcome analysis



© Copyright by International Society of Endovascular Specialists 2004